Dead Weight Tester (SMT-FM-18)

availability: in stock

Specifications:
Bourdon tube pressure gauge for pressure measurement Closed water circuit.
Transparent dial face with a view of the spring mechanism Electromagnetic flow meter.
Accurately fitting piston and cylinder of the piston manometer without seals.
Technical specifications

- +

Description

In the field of metrology, calibration describes a process for detecting deviations in a
measuring instrument compared to a reference instrument or universally accepted
standard value. This observed deviation is taken into account in the subsequent use of the
calibrated measuring instrument and adjusted if necessary. A piston manometer is
connected to a Bourdon tube pressure gauge via a pipe. Piston manometers are especially
suited to producing well-defined pressures in liquids or gases and have been used for Years
as one of the most accurate methods for calibrating pressure gauges.